Using of the white lasers at the spectral responsivity scale realisation

Автор(и)

  • Alexandr Dunaev All-Russian Research Institute for Optical and Physical measurements (VNIIOFI) Moscow, Russia, Російська Федерація

DOI:

https://doi.org/10.24027/2306-7039.1A.2017.100114

Ключові слова:

Optical and physical measurements, Spectral sensitivity, White laser

Анотація

The Russian national standards for detector calibrations at wavelengths from 250 nm up to 14000 nm are realized at the All-Russian Research Institute for Optical and Physical measurements (VNIIOFI) at 2014. It based on the CryoRad-III System from L-1 Standards and Technology, Inc. (USA). Facility for spectral responsivity measurements with using the white laser as a source of radiation is presented. Results of monochromatic power measurements in the spectral range from 0.4 to 1.6 um are presented

Біографія автора

Alexandr Dunaev, All-Russian Research Institute for Optical and Physical measurements (VNIIOFI) Moscow, Russia

Junior researcher

Посилання

S.P. Morozova et al., An absolute cryogenic radiometer for laser calibration and characterization of photodetectors, Metrologia, 1995/96, 32, 557-560.

K.D. Stock and Hofer, PTB primary standard for optical radiant power: transfer-optimized facility in the clean-room centre, Metrologia, 1995/96, 32, 545-549.

Gavrilov V.R., Dunaev A.Y., Lisyanskii B.E., Morozova S.P. et al., National Primary Standard for the Units of Absolute and Relative Spectral Sensitivity at Wavelengths from 0.25 to 14.00 μm Get 213–2014, Measurement Techniques, February 2016, Volume 58, Issue 11, pp 1200-120

##submission.downloads##

Опубліковано

2017-04-20